Page 2 D-Tec Vaporizer
August 2001 O & M Manual Part No. 1107-0622-000
1.3 Symbols
Symbol Location Facility/Rating
Control Dial Standby setting
Interlock Block Setting mark - dial marking alignment point for required setting
Interlock Block Equi-potential stud - connection for minimizing electrical potential difference
Front Panel Auditory alarm mute button
Display Panel 240 ml refill mark - indicates that an additional 240 ml of agent can be added
Display Panel Symbolizes nominal agent level display when 390 ml of agent in sump (max.)
Display Panel Symbolizes nominal agent level display when only 60 ml of agent in sump (min.)
Rear Label Mains electrical supply must be single phase sinusoidal alternating current
Top Plate Label Direction of flow
w
Various Warning
2.0 Description
2.1 General
The D-Tec Vaporizer is designed to add Suprane™ (desflurane) inhalant anesthetic vapor to the medical gases supplied
to a patient. The vaporizer must only be mounted on a Dräger compatible anesthesia machine which is equipped with
an interlocking type manifold. The vaporizer must be supplied with dry medical gas and it must be connected to a
suitable electrical mains supply. However the D-Tec vaporizer cannot be used on the Narkomed Mobile anesthesia
system.
The D-Tec Vaporizer is designed to meet UL 2601-1, IEC 601-1 and IEC 601-2-13 recommendations.
When the vaporizer is connected to the electrical supply, the dial is turned to the q (Stand-by) symbol and the green
OPERATIONAL light located on the front display panel is illuminated, the vaporizer is in the electrical stand-by condition
and is ready for use, as described in Section 5.0 Operating Instructions.