Datex-Ohmeda D-Tec Operations Instructions Page 22

  • Download
  • Add to my manuals
  • Print
  • Page
    / 90
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 21
Tec 6 Plus Vaporizer
3-6 2/16/01 1107-0641-000
Alarm Priority
The Tec 6 Plus Vaporizer has three distinct alarm priorities. For each
priority there is a different set of visual and auditory signals:
Note:
An internal processor failure may also present an intermittent red No
Output LED and alarm tone.
Figure 3-7 • Alarm priorities
Alarm Priority Visual Signal Auditory Signal
High Rapid red flashing
light
Two sets of five
audio beeps every
10 seconds
Medium Amber flashing light One set of three au-
dio beeps every 25
seconds.
Low Continuous amber
light
One audio beep
only
Vap. Condition Low Battery Low Agent No Output
Dial ON Medium
Priority
Medium
Priority
High Priority
Dial ON,
Alarm Silence
Pressed
Low Priority Audio
silenced for
120 seconds,
then audio
active
High Priority
(pressing Si-
lence has
no effect)
Dial turned
to
Low Priority Low Priority
(dial locked
until condition
corrected)
Low Priority
(dial locked
until condition
corrected)
Page view 21
1 2 ... 17 18 19 20 21 22 23 24 25 26 27 ... 89 90

Comments to this Manuals

No comments