Datex-Ohmeda D-Tec Operations Instructions Page 42

  • Download
  • Add to my manuals
  • Print
  • Page
    / 90
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 41
Tec 6 Plus Vaporizer
5-2 2/16/01 1107-0641-000
Alarm and display test
ww
ww
WARNING Only use vaporizers that are in a serviceable condition.
ww
ww
CAUTION Before connecting the vaporizer to an AC power supply,
be sure it is within the ambient operating temperature
range (see
Performance
in Chapter 9).
ww
ww
WARNING Do not use the vaporizer if during the alarm and display
test any one of the five lights and all the LCD agent level
indicator bars do not flash or the auditory alarm does not
operate for a period of approximately four seconds.
Step 1
Connect the power cord from
the vaporizer to an approved
hospital grade outlet socket.
Step 2
Verify that the alarm and
display test for a period of
approximately four seconds as
follows:
Each light and all the LCD
agent level indicator bars
on the front display panel
flash.
The auditory alarm is
activated.
Figure 1-1 • Verifying the alarm
DESFLURANE
NO OUTPUT
LOW AGENT
ALARM
BATTERY LOW
WARM-UP
OPERATIONAL
Flashes