Tec 6 Plus Vaporizer
5-10 2/16/01 1107-0641-000
Setting the dial
ww
ww
WARNING High percent dial settings combined with low gas flows
may lead to hypoxic mixtures within the breathing circuit.
Datex-Ohmeda strongly recommends the use of oxygen
monitoring.
The dial release must be operated to turn the dial from the
n setting and also to increase the dial setting to above
12%. Do not operate the dial release when turning the dial
to any other setting, it may result in overriding the 12%
stop and causing an inadvertent delivery in excess of 12%.
The vaporizer has not been calibrated at any dial setting
between
n and 1%. Do not use the vaporizer at dial
settings between
n and 1%.
Important
It is a requirement of European Standard EN 740 -
Anaesthetic Work-
stations And Their Modules
, that the gas monitoring device referred to
in
What is a Tec 6 Plus Vaporizer?
in Section 2 of this manual is in op-
erating condition, by being enabled and functioning, prior to use of the
vaporizer.
An appropriate anesthetic gas monitor must be used to monitor the
concentration of anesthetic agent vapor in the inspiratory gas when
the vaporizer is in operation, in order to provide protection against haz-
ardous output in the event of a device malfunction.
Datex-Ohmeda strongly recommends the use of anesthesia gas mon-
itoring with this equipment. Refer to local standards for mandatory
monitoring.