Tec 6 Plus Vaporizer
9-2 2/16/01 1107-0641-000
Note:
All specifications are nominal and subject to change without notice.
Calibration
Check the calibration certificate that is included with your Tec 6 Plus
Vaporizer.
ww
ww
WARNING The Tec 6 Plus Vaporizer can only be calibrated at a
Datex-Ohmeda Authorized Service Center.
The Tec 6 Plus Vaporizer is calibrated at 760 torr and at a nominal
21 °C for the following concentrations (v/v percent): 1, 2, 3, 4, 5, 6, 7,
8, 9, 10, 12, 14, 16 and 18, using O
2
as the carrier gas flowing at five
liters/minute.
Electrical supplies
.
The ~ symbol on the vaporizer rating label denotes that the external
AC power supply must be single phase sinusoidal alternating current.
Voltage requirement 100 to 120V AC 50/60 Hz or
220 to 240V AC 50/60 Hz
Power consumption 250 VA nominal
Current input Maximum current requirement is 2.0 A