Tec 6 Plus Vaporizer
3-10 2/16/01 1107-0641-000
No output
This alarm indicates that the vaporizer has shut down and is no longer
delivering vapor.
The NO OUTPUT alarm is activated if the dial is in the ON position and
any one of the following occurs:
The agent level decreases to below 20 ml.
The vaporizer is tilted for more than 10 seconds.
A power failure of longer than 10 seconds duration occurs.
An internal malfunction is detected.
Figure 3-11 • No output
If the NO OUTPUT alarm occurs while the vaporizer is in use, the va-
porizer dial must be turned to .
This action de-escalates the alarm to a low priority and illuminates the
amber light continuously. Either an alternative vaporizer or an alterna-
tive method of anesthesia must be used.
If the NO OUTPUT alarm is activated as a result of a power failure and
the dial is at when the power supply is restored, the alarm and dis-
play test is activated and the vaporizer returns to either the WARM-UP
or the OPERATIONAL mode.
If the dial is not at when the power supply is restored, the NO OUT-
PUT alarm continues and can only be cancelled by turning the dial
to . In this condition, the alarm and display test is not activated and
the vaporizer returns to either the WARM-UP or the OPERATIONAL
mode.
AA.43.042
DESFLURANE
NO OUTPUT
LOW AGENT
ALARM
BATTERY LOW
WARM-UP
OPERATIONAL